Flanged ECT Phantom

Flanged ECT Phantom

Phantom for high resolution NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters).

Among other things, it can be used to evaluate: COR error, non-uniformity artifacts, changes of ROR on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

 

Features

Complies with:

·     NEMA Standards Publication (NU 1-2001) Performance measurements of Scintillation Cameras

·    AAPM Report No. 9, Computer Aided Scintillation Camera Acceptance Testing

·    AAPM Report No. 22, Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control

·    ACR-SNM [Res. 5-2011) technical standard for diagnostic procedures using radiopharmaceuticals

 

Specifications

·    Inside main cylinder diameter: 190 mm

·    Inside main cylinder height: 195 mm

·    Main cylinder wall thickness: 5 mm

·    Rod diameters: 6, 8, 10, 12, 13 and 15 mm

·    Height of rods: 90 mm

·    Solid sphere diameters: 12.7, 15.9, 19.1, 25.4, 31.8, 38 mm

·    Height of the center of the spheres from the base plate: 130 mm

·    Refillable thin walled cylinders, diameters: 8, 12, 16 and 25 mm

·    Water filled, Air filled and PTFE solid cylinder diameter: 25 mm

·    Cylinders height: 38 mm

Score: 5 from 5. Sum 1 Vote
Replying to: [Cancel]
Remaining characters: 500
Captcha  
03/01/2024 | Types of phantom | 1669568487 | Hit: 11 | 1704291454